ProLab 1000

Operation

47

ba75576e05

11/2013

4.5.7

Calibrating

Make sure the Buffer setting is correctly set in the sensor menu (see
page 38).

Use any one to five buffer solutions of the selected buffer set in
ascending or descending order.

The NIST/DIN calibration is described below. With other buffer sets,
other nominal buffer values are displayed. Apart from that, the
procedure is identical.

1

Press <MODE> to select the measured parameter pH or mV in
the measured value display.

2

Start the calibration with <CAL>.
The calibration display appears.

3

Immerse the electrode in buffer solution 1.

4

If the Set temperature menu item appears, measure and enter
the temperature of the buffer manually (measurement without
temperature sensor).

5

Select Continue with <

▲><▼> and press <MENU/OK>. The

buffer is measured.
The measured value is checked for stability (Stability control).

[i] Buffer series NIST/DIN

[i] Immerse sensor in buffer 1

Continue

pH <CAL>

[i] Buffer value = 6.865

[i] U = 3.0 mV

[i] Temperature =

24.8 °C

Terminate stability control

Recognized nominal buffer
value (referring to 25 °C)

Current electrode voltage

Current temperature value

pH <CAL>