4.4.4

Automatic Stability control

The automatic Stability control function continuously checks the
stability of the measurement signal. The stability has a considerable
impact on the reproducibility of measured values.

You can activate or switch off the automatic Stability control function
(see page 31).

The automatic Stability control function is carried out:

z as soon as the measured value is outside the allowed stability range
z when you switch over between the measured parameters with

<MODE>.

In this case, the measured parameter flashes on the display.