8 measurements with dead stop function, Measurements with dead stop function – Xylem ProLab 1000 BENCH TOP pH METER User Manual

Page 51

Advertising
background image

ProLab 1000

Operation

51

ba75576e04

01/2011

4.5.8

Measurements with dead stop function

Measurements with dead stop function are normally used for measured
value logging during a manual dead stop titration (e.g. Karl Fischer
titration).

A voltage is present at the dead stop sockets. The voltage value is con-
stant before the end point of the titration is reached.
Even with a small excess of titration solution, a very strong voltage drop
occurs.
Thus the end point of the titration is reached.

Preparatory activities

1

Connect the pH socket and dead stop socket with the dead
stop cable.

2

Connect a double Pt electrode, e.g. type Pt 1400, to the dead
stop sockets of the meter.

3

Immerse the double Pt electrode in the test sample.

4

On the meter, switch to the mV display with <MODE>.

Double Pt-electrode

Dead stop cable