3 general errors, General errors – Xylem ProLab 1000 BENCH TOP pH METER User Manual

Page 82

Advertising
background image

What to do if...

ProLab 1000

82

ba75576e04

01/2011

Warning [TempErr]

Warning [ISEErr]

6.3

General errors

Meter does not react

to keystroke

Time is lost

Cause

Remedy

– Temperature difference

between measurement and
calibration greater than 2 K.

– Adjust the temperature of

the test sample

Cause

Remedy

– Electrode voltage outside

calibrated range

– Recalibrate

Cause

Remedy

– Operating condition undefined

or EMC load unallowed

– Processor reset:

Press the <On/Off> and
<SC> key simultaneously.

Cause

Remedy

– Buffer batteries empty

– Exchange buffer batteries

(see page 77)