4 moving patient on tilted stretcher, 5 ambulance transport, 8 changing air sources – Physio-Control LUCAS User Manual

Page 21: Hanging, Sources

Advertising
background image

100057-00 F, ©J

OLIFE

AB 2007

21

LUCAS

TM

C

HEST

C

OMPRESSION

S

YSTEM

, I

NSTRUCTIONS

FOR

U

SE

3.7.4 M

OVING

PATIENT

ON

TILTED

STRETCHER

In situations where conditions do not allow the

chest to remain horizontal:

1.

Turn the

ON/OFF

knob to

Lock

to

pause

compressions (Figure

1(

).

2. As soon as the patient is in horizontal position,

check the position of the suction cup and

adjust if necessary (see Section 3.4).

3. Turn the

ON/OFF

knob to

Active

.

3.7.5 A

MBULANCE

TRANSPORT

Make sure that the patient is fixed when using

LUCAS Chest Compression System during

ambulance transport. The ambulance cot should
have four-point restraints for the patient’s upper

body in addition to the two lower-body restraints.

3.8 C

HANGING

AIR

SOURCES

WARNING – INTERRUPTED COMPRESSIONS

There may be excessive changeover time if pressure

regulators have to be mounted.

Always use reserve cylinders with pre-fitted

pressure regulators.

To change air source:

1.

Set the

ON/OFF

knob to

Adjust

.

2. Close the valve on the used air cylinder.
3. Disconnect the air hose from the air source.
4. Open the valve on the new air source.
5. Attach the air hose to the new air source.
6. Adjust the height of the suction cup on the

patient’s chest, as described previously on
page 17.

7. Turn the

ON/OFF

knob to

Lock

.

8. Turn the

ON/OFF

knob to

Active

to

continue compressions.

1(

LOCK