Appendix A

W

EEKLY

CHECK

AND

CHECK

AFTER

USE

, LUCAS

Use this protocol for the weekly check on the device and the check after use. Store the completed checklist.

Copying of this sheet is permitted.

If any error cannot be corrected take the device out of operation and indicate the problem in the “Serious problem”
column. Then immediately contact your distributor for advice or to order a new product (www.jolife.com).

When carrying out these checks always use air from a central air supply or separate air cylinder. Do not use up

air from a air cylinder intended for use in emergency situations.

Date:

Action

Approved

signature

Adjustments

carried out

Serious problem

Check that the device is clean.

Check that the air hose is not cracked and fits

tightly to LUCAS.

Check that the connector(s) is/are present and

undamaged.

Check that the

ON/OFF

knob is in the

Adjust

position.

Pull the release rings so that the claw locks are

open for assembly. Attach the upper part to the

back plate.

Check that the compression mechanism can be

adjusted from the uppermost to the lowermost

position without any jarring or sluggishness.

Attach the air hose to a air supply. Switch the

device on for 1–2 minutes. Check that it is

operating normally. Calculate the number of

compressions per 60 seconds. 100 compressions/

min ± 10 is acceptable.