6 Technical

specification

6.1 O

PERATION

All pressure regulators supplied by your
distributor are approved for use with LUCAS.

Power source

Air

Air

consumption

13.7 gallons/minute (52 litres/minute)

Air hose

Permanently attached, 10.8 feet (3.3 m)

in length, with unique male connector.

Approved air

sources for use

with LUCAS

Pressure regulators or wall outlets for air

(medical grade/breathing quality) should

meet the following specifications:
1. Nominal supply pressure: 46-87 PSI

(3.17-6 bar).

2. Minimum required free flow rate in

relation to nominal pressure

according to Figure

2!

. Example

shown: Nominal supply pressure 66.7

PSI (4.6 bar).

3. Pressure drop at minimum free flow

rate obtained from Figure

2!

, must

not exceed the value obtained from

Figure

2@

. The shaded area in

Figure

2#

illustrates an unacceptably

high pressure drop for nominal supply

pressure 66.7 PSI (4.6 bar) and above.

The performance of the air supply

should be verified according to the

procedures in Section 6.1.1 and 6.1.2. If

you do not know how to do this, please

contact a clinical engineer or your

distributor, for help.