Event review screens, Troubleshooting tips – Physio-Control TrueCPR User Manual

Page 16

Advertising
background image

12

TrueCPR Device Instructions for Use

Troubleshooting Tips

Event Review Screens

Two event review screens display data for the most recent device use. To display
the event review screens, start with the device turned off. Press the

(AIRWAY)

and

(POWER) buttons simultaneously to turn the device on in Event Review

mode. To switch between the screens, press the

(MUTE) button.

Troubleshooting Tips

These troubleshooting tips provide immediate corrective actions for possible
problems. If a problem persists, contact your local Physio-Control representative
for assistance. In the USA, call Technical Support at 800.442.1142.

Percent of compressions with depth > 5 cm
(2 in).

Average compression rate (compressions per
minute).

Total event duration.

Percent of total event with compressions.

OBSERVATION

CORRECTIVE ACTION

No audible metronome.

• Use compression rate meter for guidance.

Device does not function.

• Only use device within specified operating

temperature range of 0°–40°C (32°–104°F).

• Remove batteries, wait 30 seconds, and

then re-install batteries.

• Perform CPR unaided.

Display is visible, but does not
provide compression feedback.

• Turn device off and back on. Do not apply

pressure to chest pad until calibration is
complete and feedback screen appears.

• Perform CPR unaided.

Alert indicator persists.

• Turn device off and back on. If indicator

persists, perform CPR unaided.

• Contact Technical Support or your local

Physio-Control representative for
assistance.