Table 3 Guidance and Manufacturer’s Declaration - Electromagnetic Immunity

The TrueCPR device, model 80596-000003, is intended for use in the electromagnetic
environment specified below. The customer or user should ensure that the TrueCPR
device is used in such an environment.

Immunity Test

IEC 60601

Test Level

Compliance

Level

Electromagnetic Environment -

Guidance

Portable and mobile RF
communications equipment should be
used no closer to any part of the
TrueCPR device, including cables, than
the recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.

Recommended separation distance

Conducted RF
IEC 61000-4-6

3 Vrms
150 kHz to
80 MHz

3 Vrms

Radiated RF
IEC 61000-4-3

3 V/m
80 MHz to
2.5 GHz

3 V/m


80 MHz to 800 MHz


800 MHz to 2.5 GHz

Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,

a

should be

less than the compliance level in each
frequency range.

b

Interference may occur in the vicinity of
equipment marked with the following
symbol:

Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.

d

1.2 P

=

d

1.2 P

=

d

2.3 P

=