Troubleshooting Tips for Data Management

Cleaning the Device

The TrueCPR device should be cleaned and inspected after each use.

1

Clean the device with a sponge or cloth that has been dampened with one of
the cleaning agents shown below:

Quaternary ammonium compounds

Hydrogen peroxide, 3% solution

Sodium dichloroisocyanurate (NaDCC), 3000 ppm solution

Chlorine bleach, 1:10 dilution with water

2

Inspect the device and cable for wear or damage. Remove from service if
damaged.

3

After cleaning, wipe off solution with a clean, damp cloth.

IMPORTANT! Do not immerse or soak the chest pad or back pad.

Storing the Device

Place chest pad onto back pad. Store in carrying bag so that battery readiness
indicator is visible.

OBSERVATION

CORRECTIVE ACTION

Feedback screen appears on
TrueCPR device and
metronome sounds.

• Confirm USB cable is properly connected.

• Confirm computer is on.

• Replace USB cable.

• Use recommended USB cable (see Step 1

on page 14).

Data Transfer Failed screen
appears.

• Turn TrueCPR device off and check

connections. Then turn device on and
attempt data transfer again.

• If problem persists, contact Technical

Support.

CAUTION

POSSIBLE DEVICE DAMAGE: If the TrueCPR device will be stored for an
extended period of time, remove the batteries to prevent possible damage
due to battery leakage.