Electromagnetic emissions, Essential performance, Electromagnetic immunity – Physio-Control TrueCPR User Manual

Page 25: Electromagnetic compatibility guidance, English

Advertising
background image

©2012 Physio-Control, Inc.

19

English

Electromagnetic Compatibility Guidance

Electromagnetic Emissions

The TrueCPR device emits a continuous wave signal at 12 ± 0.1 kHz. The maximum
Effective Isotropic Radiated Power (EIRP) and Effective Radiated Power (ERP) levels are
-40 dBm, measured at a distance of 3 m (10 ft).

Essential Performance

The TrueCPR device maintains safe and effective performance (accurate depth and recoil)
when operated in the electromagnetic environment specified in Table 2 through Table 4 .

Electromagnetic Immunity

Table 1 Guidance and Manufacturer’s Declaration - Electromagnetic Emissions

The TrueCPR coaching device, model 80596-000001, is intended for use in the
electromagnetic environment specified below. The customer or user should ensure that
the TrueCPR device is used in such an environment.

Emissions Test

Compliance

Electromagnetic Environment - Guidance

RF emissions
CISPR 11

Group 1

The TrueCPR device uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference
in nearby electronic equipment.

RF emissions
CISPR 11

Class B

The TrueCPR device is suitable for use in all
establishments including domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
The TrueCPR device is battery-powered.

Harmonic emissions
IEC 61000-3-2

Not applicable

Voltage fluctuations/
flicker emissions
IEC 61000-3-3

Not applicable

Table 2 Guidance and Manufacturer’s Declaration - Electromagnetic Immunity

The TrueCPR device, model 80596-000001, is intended for use in the electromagnetic
environment specified below. The customer or user should ensure that the TrueCPR
device is used in such an environment.

Immunity Test

IEC 60601

Test Level

Compliance

Level

Electromagnetic Environment -

Guidance

Electrostatic
discharge (ESD)
IEC 61000-4-2

±6 kV contact
±8 kV air

±6 kV contact
±8 kV air

Floors should be wood,
concrete, or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.

Electrical fast
transient/burst
IEC 61000-4-4

±2 kV for power
supply lines
±1 kV for input/
output lines

Not applicable

The TrueCPR device is battery-
powered.

3308456-900.book Page 19 Wednesday, October 24, 2012 9:51 AM