Measurement Throughput Summary

Table 14. Typical cycle time

1

(ms) for measurement completion

Number of Points

201 401 801 1601

16,001

Start 8 GHz, stop 18 GHz, 600 kHz IF bandwidth
Uncorrected

21.148 21.743 23.01 25.198 54.836

4-Port cal

74.597

82.296

296.5

307.75 538.646

Start 300 kHz, stop 10 GHz, 600 kHz IF bandwidth
Uncorrected 19.814

22.801

24.973

29.01

67.733

4-Port

cal

69.752 85.111 100.125 129.347 480.711

Start 300 kHz, stop 20 GHz, 600 kHz IF bandwidth
Uncorrected 32.575

34.7

39.237

43.155

69.625

4-Port

cal

121.254 133.626 157.506 179.223 487.779

Start 8 GHz, stop 18 GHz, 100 kHz IF bandwidth
Uncorrected

38.083 51.816 55.488 56.36 184.154

4-Port

cal

143.271 201.814 215.056 230.133 934.161

Start 300 kHz, stop 10 GHz, 100 kHz IF bandwidth
Uncorrected

37.03 42.532 45.122 46.729 198.683

4-Port

cal

137.431 162.37 194.13 192.182 906.768

Start 300 kHz, stop 20 GHz, 100 kHz IF bandwidth
Uncorrected

44.98 69.408 87.161 92.475 198.792

4-Port

cal

169.041 268.877 343.898 369.526 914.963

Start 8 GHz, stop 18 GHz, 50 kHz IF bandwidth
Uncorrected 42.171

70.09

88.702

90.981

371.611

4-Port

cal

157.107 271.791 351.517 368.02 1532.609

Start 300 kHz, stop 10 GHz, 50 kHz IF bandwidth
Uncorrected 43.713

61.41

66.878

69.373

385.04

4-Port cal

163.58 238.267 259.687 279.816 1580.761

Start 300 kHz, stop 20 GHz, 50 kHz IF bandwidth
Uncorrected

48.673 80.798 124.605 147.303 388.46

4-Port

cal

184.429 313.392 493.142 587.548 1587.839