Table 15. Cycle Time vs IF Bandwidth

Applies to the preset condition (201 points, correction off) except for the following changes:
• CF = 10 GHz
• Span = 100 MHz
• Display off (add 21 ms for display on)

Description Typical

performance

IF Bandwidth (Hz)

Cycle time (ms)

1

Trace noise (dB rms)

600,000 3.13

0.00544

360,000 3.21

0.00602

280,000 3.17

0.00321

200,000 3.17

0.00259

150,000 3.19

0.00207

100,000 4.05

0.00155

70,000 4.99 0.00144
50,000 6.41 0.00121
30,000 8.78 0.00094
20,000 12.07

0.00080

15,000 14.91

0.00069

10,000 26.02

0.00052

7000 34.54 0.00047
5000 45.87 0.00044
3000 69.91 0.00032
2000 99.69 0.00029
1500 128.18

0.00010

1000 215.62

0.00009

700 291.58 0.00006
500 397.36 0.00007
300 633.86 0.00000
200 930.15 0.00000
100 1824.19

0.00000

30 6001.70 0.00000
10 17899.79

0.00000

1 178391.58

0.00000

1. Cycle time includes sweep and retrace time.

Advertising