Atec Agilent-N5230A User Manual

Page 28

Advertising
background image

28

Table 16. Cycle time vs number of points

Applies to the preset condition (correction off) except for the following changes:
• CF = 10 GHz
• Span = 100 MHz
• Display off (add 21 ms for display on)

IF Bandwidth (Hz)

Number of points

Cycle time (ms)

1

30,000 3

2.88

11

3.50

51

3.91

101

5.29

201

8.75

401

15.66

801

29.46

1,601

57.73

6,401

221.04

16,001

549.71

100,000 3

2.87

11

2.82

51

2.86

101

2.96

201

4.02

401

6.23

801

10.65

1,601

19.49

6,401

70.96

16,001

173.78

600,000 3

2.84

11

2.84

51

2.87

101

3.03

201

3.14

401

3.51

801

4.22

1,601

6.22

6,401

19.35

16,001

45.12

1. Cycle time includes sweep and retrace time.

Table 17. Data transfer time (ms)

Number

of

points

201

401

1601

16,001

SCPI over GPIB (program executed on external PC)
32-bit floating point

6

10

33

313

64-bit floating point

10

18

65

622

ASCII

53

105

406

4032

CPI over SIU/LAN or TCP/IP socket (program executed in the analyzer)
32-bit floating point

1

2

2.5

7

64-bit floating point

2

2

3

10

ASCII

11

20

73

720

COM (program executed in the analyzer)
32-bit floating point

< 0.2

0.2

0.3

0.9

Variant

type

0.6 1 3.2 32

DCOM over LAN (program executed on external PC)
32-bit floating point

< 0.8

1

1.6

7.5

Variant

type

1.9 3 8.9 82

Note: Specifications for recall and sweep speed are not provided for the N5230A analyzers.