ESG Family Signal Generators

Adjustments

ALC Modulation Flatness Calibration

Calibration Guide

5-39

ALC Modulation Flatness Calibration

Description

This test calibrates the ALC REF DAC corrections necessary to maintain power flatness in
the ALC open-loop mode of operation. The nominal ALC REF DAC settings are adjusted to
re-align the ALC open-loop power with the desired level. The resulting values are stored as
calibration arrays in the signal generator’s firmware.

Required Test Equipment

• HP/Agilent 438A Power Meter

• HP/Agilent 8482A Power Sensor

Procedure

Figure 5-33

ALC Modulation Flatness Calibration Setup

1. Connect the equipment as shown above.

2. Preset all of the equipment.

3. Follow the instructions as they appear on the controller’s display.