Performance Tests

ESG Family Signal Generators

5. AM Frequency Response

4-14

Calibration Guide

5. AM Frequency Response

This automated test verifies the AM frequency response specifications. The equipment
measures the variations in modulation depth due to changes in the applied AM rate; dc to
10 kHz. The response is relative to a 1 kHz rate set at the test depth, and expressed in dB.
Each frequency is tested with the following sequence:

1. A reference signal is measured.

2. The set depth is measured for each applied rate change.

3. Each depth measurement is compared to the reference signal depth.

The comparison is a calculation of the difference in the depths relative to the reference
rate. The relative value is the AM Frequency Response.

Recommended Equipment

• HP/Agilent 8902A Measuring Receiver Option 010

• HP/Agilent 8904A Function Generator

• HP/Agilent 8663A Signal Generator

• MD/MDC-164 Mixer (0.5 to 9 GHz)

Equipment Setups

Figure 4-12 Frequency Response for AM

1300 MHz Test Setup