Table 2-3, As appropriate – Atec Agilent-8563E User Manual

Page 15

Advertising
background image

ESG Family Signal Generators

Operation Verification

Verification Procedures

Calibration Guide

2-13

Table 2-3 Power Level Accuracy Setup 1 (Signal Generators with Option 1E6)

Frequency

Setting

Power Level

Setting

(dBm)

Lower Limit

(dBm)

Measured

Power

(dBm)

Upper Limit

(dBm)

277 kHz

+9

8.2

9.8

+3

2.5

3.5

0

0.5

0.5

5

5.5

4.5

15

15.5

14.5

2.516 MHz

+9

8.2

9.8

+3

2.5

3.5

0

0.5

0.5

5

5.5

4.5

15

15.5

14.5

270.1 MHz

+9

8.2

9.8

+3

2.5

3.5

0

0.5

0.5

5

5.5

4.5

15

−1

5.5

14.5

510.1 MHz

+9

8.2

9.8

+3

2.5

3.5

0

0.5

0.5

5

5.5

4.5

15

15.5

14.5

990.1 MHz

+9

8.2

9.8

+3

2.5

3.5

0

0.5

0.5

5

5.5

4.5

15

15.5

14.5

1350.1 MHz

+6

5.2

6.8

+3

2.5

3.5

0

0.5

0.5