Damping mats (optional), Selection of damping mats, Quantity for pipe damping mat - type a – Atec Flexim-G601 User Manual

Page 24: Length of pipe damping mat - type b

Advertising
background image

24

TSFLUXUS_G601V1-5-1US_Lus, 2014-02-25

FLUXUS® G601

Technical Specification

Damping Mats (optional)

Damping mats will be used for the gas measurement to reduce acoustic noise influences on the measure-

ment.
Transducer damping mats will be installed below the transducers.
Pipe damping mats will be installed at reflection points, e.g. flange, weld.

Selection of Damping Mats

type

description

outer pipe

diameter

dimensions

l x b x h

transducer

frequency

techni-

cal type

ambient

temperature

remark

in

in

G H K M P

°F

transducer damping mat
D

for temporary installation
(multiple use), fixed with
coupling compound

< 3.1

17.72 x 4.53 x 0.02 -

-

-

x

x D20S3

-13 to +140

≥ 3.1

35.43 x 9.06 x 0.02 -

-

x

x

- D20S2

35.43 x 9.06 x 0.05 x

x

-

-

- D50S2

pipe damping mat
A

for temporary installation
(multiple use), fixed with
coupling compound

< 11.8

11.81 x 4.53 x 0.02 x

x

x

x

x A20S4

-13 to +140

for quantity see
table below

B

self-adhesive

≥ 11.8

l x 3.94 x 0.04

x

x

x

x

x B35R2

-31 to +122

l - see table below

Quantity for Pipe Damping Mat - type A

(depending on the outer pipe diameter)

outer pipe diameter D

transducer frequency

in

G, H

K, M, P

3.9

12

6

7.9

24

12

11.8

32

16

Length of Pipe Damping Mat - type B

(length l depending on transducer frequency and outer pipe diameter)

outer pipe diameter D

transducer frequency

G, H

K, M, P

in

ft

ft

11.8

39

19

19.7

104

52

39.4

413

206

b

l

D

transducer damping mat

pipe damping mat

reflect arrangement

diagonal arrangement/direct mode

D = outer pipe diameter

reflection point