Model 2290-5 High Voltage Power Supply Characteristics

Specifications and characteristics are subject to change without notice.
Page 2 of 3

SPEC-2290-5 Rev. A / December 2013

CHARACTERISTICS

Voltage range:

Discharge time (full load)

5, 6

:

1 second for 5000 V DC to 1 V DC – No filter

3 seconds for 3000 V DC to 1 V DC – Filter 1

4 seconds for 5000 V DC to 1 V DC – Filter 2

Discharge time (no load)

5

:

<12 seconds (to <50 V DC) – No filter

<30 seconds (to <50 V DC) – Filter 1 or Filter 2

Output stored charge: <0.9 mC maximum

Settling time

5, 6

: From 0 to programmed voltage; to within 99.9% of final value within 3 seconds

Recovery time

5, 6

: 120 ms for 40% step change in load current (typical)


Current limit and trip range

Filter

0 mA to 5.25 mA

No filter or Filter 1

0 mA to 3.25 mA

Filter 2

Current set accuracy

7

: ±(0.01% of setting + 2.5 µA)

Current resolution: 1 µA

Current display accuracy: ±1 µA, typical (±2 µA, maximum)

Stability: ±0.01% per hour; <0.03% per eight hours

Temperature drift: 50 ppm/ºC, 0º to 40º C, typical

Protection: Arc and short circuit protected; programmable voltage and current limits and current trip

Monitor outputs

Output scale: 0 to +10 V to full scale

Current rating: 10 mA maximum

Output impedance:

<1 Ω

Accuracy: ±0.2% of full scale

with a 100 kΩ load, minimum

Update rate: 8 Hz

7

Add 2.5 µA offset when Filter 1 or Filter 2 is enabled.

Advertising