Model 2290-5 High Voltage Power Supply Characteristics

Specifications are subject to change without notice.
SPEC-2290-5 Rev. A / December 2013

Page 3 of 3

External voltage set

Input scale: 0 to +10 V for 0 to full scale

Input impdance: 1 M

Accuracy: ±0.2% of full scale

Update rate: 16 Hz

Output slew rate

5, 6

: <(Rise time + 0.3 seconds) for 0 to full range under full load

GENERAL:

Input power: 55 watts

2290-5 Input voltage: 120 V ±10%, 50 or 60 Hz

2290E-5 Input voltage: 240 V ±10%, 50 or 60 Hz

2290J-5 Input voltage: 100 V ±10%, 50 or 60 Hz

Rear panel connectors:

SHV male (Kings type 1704-1 or equivalent)

Output high-voltage connector GPIB connector

BNC Connector (two): Voltage set/Voltage monitor; Current monitor

Toggle switch: Voltage setting or Voltage monitor

High-voltage safety interlock:

Connector: 3-pin press-in connector, 3M part number: 37103-A165-00E-MB

Pin 1: 5 V nominal out, 1.5 mA maximum out

Pin 2: Input: High-voltage output enabled: 3 – 24 V DC

High-voltage output disabled: <1.2 V DC or open connection

Pin 3: Chassis ground through a 100

resistor

Interface protocol: IEEE-488.1

Operating environment: 0º C to 40º C; non-condensing

Dimensions: 89 mm high x 206 mm wide x 406 mm deep (3.5 in x 8.1 in x 16 in)

Weight: 5.5 kg (12 pounds)

Safety: Conformance to European Union low voltage directive

Warranty: One year

Warm-up time: One hour