B&K Precision 879B - Manual User Manual

Page 92

Advertising
background image

90

Frequency = 1 kHz

Range

Max

Display

Lx

Accuracy

DF(Dx

<0.5)

Measurement

Mode

100 H

100.00 H

1.5% + 3

digits

1.5% +

50 digits

Parallel

40 H

39.999 H

0.7% + 2

digits

0.7% +

50 digits

Parallel

4000 mH

3999.9 mH

0.7% + 2

digits

0.7% +

50 digits

---

400 mH

399.99 mH

0.5% + 1

digits

0.5% +

50 digits

Series

40 mH

39.999 mH

0.6% + 2

digits

0.6% +

50 digits

Series

4000 μH

3999.9 μH

0.9% + 2

digits

0.9% +

50 digits

Series

400 μH

399.99 μH

2.8% + 3

digits

2.8% +

50 digits

Series

Frequency = 10 kHz*

Range

Max

Display

Lx

Accuracy

DF(Dx

<0.5)

Measurement

Mode

1000 mH

1000.0 mH

1.5% + 3

digits

1.5% +

50 digits

Parallel

400 mH

399.99 mH

0.7% + 2

digits

0.7% +

50 digits

---

40 mH

39.999 mH

0.5% + 1

digits

0.5% +

50 digits

Series

4000 μH

3999.9 μH

0.6% + 2

digits

0.6% +

50 digits

Series

Advertising
This manual is related to the following products: