B&K Precision 2658A - Manual User Manual

Page 79

Advertising
background image

73

* When the remote control is ON, “REMOTE” is displayed in the operating

information area on the screen. ( Refer to “4. Explanation of Screen” for the details.)


51)

Single sweep

Command : CAPT * It sweeps only once and will be in a HOLD state.

52)

Setting offset level

Command : OFFSET* * * * *

(* * * * *: -50.0 to 50.0 [0.1 step, unit: dB] )

Example: OFFSET-25.6

Query: OFFSET?

53)

Setting input impedance

Command : IMP* *

(* * : 50 / 75)

Example: IMP75

Query: IMP?

* When the input impedance is selected, the offset level is set as described above.

54)

Clearing spectrum data and parameter

Command : MCLR* * *

( * * *: ALL / 000 to 199)

Example: MCLRWALL

* For internal memory only.


55)

Setting character for label

Command : LBL* * *

( * * *: The number of characters is 16 or less)

Example: LBLsample_spectrum

Query: LBL?

56)

Setting clock

Command : CLC * * * * * * * *

( * * * * * * * * : aabbccdd ; aa: two last digits of year, bb: month (01 to 12), cc: hour

(00 to 23), dd: minute (00 to 59)

Example: CLC0902111103

Query: CLC?

Command

Offset level

50

Offset level is set to 0dB.

75

Offset level is set to 5.7dB.

Command

Cleared data

ALL

All of spectrum data

000

Spectrum data of 000

· · ·

· · ·

199

Spectrum data of 199