Brooks, Ar-mite, Mt3750c – Brooks Instrument 3750 User Manual

Page 10

Advertising
background image

1-4

Installation and Operation Manual

X-VA-MT3750C-eng

Part Number: 541B063AAG

April, 2014

Brooks

®

Ar-Mite

TM

MT3750C

Section 1 Introduction

1-3 Optional Equipment

Standard Needle Valves
The standard needle or cartridge valves can be supplied integrally
mounted to the inlet fitting of the instrument. The optional NRS

TM

needle

valve, also integrally mounted to the inlet fitting, provides a greater number
of turns affording precision control. For more details on the needle or
cartridge valves go to our website: BrooksInstrument.com,
select Documentation,Precision Valves and Flow Controllers,
Brooks-Line IV

TM

, CART valves or NRS valves.

Limit Switches
Inductive Alarm Switch
One or two electronic limit switches type SJ2-N can be installed in the
indicator housing to allow initiation of signaling or switching functions on a
preset flow value being reached. The SJ2-N limit switch operates as a slot
initiator that is inductively actuated by a cam mounted to the pointer. Any
flow value can be used for setting the limit value by sliding the switch along
the slot in the mounting plate for the initiators. Minimum setting distance
between two limit switches is approximately 50% of the scale range.
(Refer to Figure 1-1)

Power supply

8 Vdc (Max. 15.5 Vdc)

Current consumption

active area clear: > 3 mA

Current consumption

active area obscured: < 1 mA

Self inductance

29 μH

Self capacitance

20 nF

Max Temp:

158°F(70°C)

Inductive Alarm
Electrical Classification
Intrinsically Safe:
ATEX: KEMA 02ATEX1126

II 2 G Ex ia IIC T6

per EN 60079-0: 2006 EN 60079-11: 2007
EN 61241-0: 2006 EN 61241-11: 2006

II 2 D Ex iaD 21 IP65 T 75

O

C

IECEx KEM 09.0046
Ex ia IIC T6 Gb/Ex ia IIIC T 75

O

C Db IP65

per IEC 60079-0: 2007-10 IEC 60079-11: 2006
IEC 61241-11: 2005

CSA: (USA and Canada) 1379260

Class I, II, III, Div. 1, Groups A thru G, T6

per CSA -157:1992 ; UL913:2002

Class I, Zone 0, Zone 1 AEx ia IIC, T6

per ANSI/ UL 2279: 1996

Ex ia IIC

per CSA E79-0:2002; CSA E79-11:2002

Advertising