Brooks, Ar-mite, Mt3750c – Brooks Instrument 3750 User Manual

Page 14

Advertising
background image

1-8

Installation and Operation Manual

X-VA-MT3750C-eng

Part Number: 541B063AAG

April, 2014

Brooks

®

Ar-Mite

TM

MT3750C

Section 1 Introduction

Intrinsic Safety:
Functional parameters

Power Supply 8-28 Vdc, Imax= 22 mA

Entity Parameters:

Ui = 30 V DC, Ii = 100 mA Pi = 750 mW
Ci = 0uF Li = 1.8 mH

Ambient Temperature: -30

o

C < Tamb < 65

o

C

ATEX: KEMA 01ATEX1033

II 2 G Ex ia IIC T6
II 2 D Ex iaD 21 IP66/IP67 T70°C
II 2 D Ex tD A21 IP66/IP67 T70°C

per EN 60079-0: 2006 EN 60079-11: 2007
EN 61241-0: 2006 EN 61241-11: 2006

CSA: (US and Canada)

IS Class I, II, III, Div.1, Groups A, thru G, T6

per UL913:2002; CSA -157:1992

Class I, Zone 1 AEx ia IIC T6

per UL 2279:1996

Ex ia IIC T6

per CSA E79-0:2002; CSA E79-11: 2002

NEPSI: (China) GYJ11.1637

Ex ia IIC T6 Gb Ex iaD 21 T 70°C

per GB 3836.1/4-2010, IEC 61241-0:2004, GB 12476.4-2010

IECEx: KEM 06.0037

EX ia IIC T6
per IEC 60079-0: 2004; IEC 60079-11:1999

GOST-R 0813793
Approval POCC NL.

06.B01220

Advertising