Brooks, Ar-mite, Mt3750c – Brooks Instrument 3750 User Manual

Page 33

Advertising
background image

3-1

Installation and Operation Manual
X-VA-MT3750C-eng
Part Number: 541B063AAG
April, 2014

Brooks

®

Ar-Mite

TM

MT3750C

Section 3 Operation

3-1 Operating Procedure

After the flowmeter has been properly installed in the process, it is ready
for operation. When initiating flow, slowly open the valve to avoid a flow
surge.Bypass is a help in bringing the flow on smoothly. Avoid starting a
pump to supply the flowmeter without the use of a valve upstream of the
flowmeter.

3-2 Operation of Inductive Alarm

A. Start-up the meter as described in Section 3-1
B. To modify the alarm set points, remove the side cover of the indicator

housing by removing the four screws.

C. Set the alarm position by loosening the screw mounting the switch to

the plate.

D. Hold the pointer to the desired alarm flow rate.
E. Move switch assembly until the metal initiator, attached to the pointer,

is not inserted into the switch.

F.

Tighten the screw mounting the switch to the plate.

G. Replace the indicator housing cover and gasket and secure with the

four screws.

3-3 Operation of Transmitter

A.

Start-up meter as described in Section 3-1.

B.

Programming is performed prior to shipment. The transmitter is
preprogrammed prior to shipment for the following parameters based
on the order information, meter configuration and application. If
parameters are not specified in the customer purchase order, the
defaults inherent to the electronics will be as shown in (parenthesis).

Pre-Programmed Parameters

Serial Number
Flow Rate units of measure (gpm)
Low Flow cutoff (5% FS flow)
Analog Output LoRange

(4 mA @ 0% flow)

Analog Output HiRange

(20 mA @ 100% flow)