Model gt1000 series – Brooks Instrument GT1000 User Manual

Page 24

Advertising
background image

3-2

Section 3 Operation

Model GT1000 Series

Installation and Operation Manual

X-VA-GT1000-eng

Part Number: 541B078AAG

March, 2014

Figure 3-1 Typical Flowmeter Piping Configuration

A - Inlet Valve

B - Outlet Valve

C - Bypass Valve

D - Control Valve

E - Drain Valve

HORIZONTAL

LINE

D

A

E

C

VERTICAL

LINE

B

B

A

E

FLOWMETER

C

D

FLOWMETER

To initiate flow through a flowmeter using bypass piping, refer to Figure 3-1.

1. Close flowmeter inlet valve (A) and outlet valve(B).

2. Fully open bypass valve (C) and slightly open control valve (D).

3. Initiate process flow. When flow has stabilized, fully open outlet valve (B),

then slowly open inlet valve (A) .

4. Close bypass valve (C).

5. Regulate process flow using control valve (D).

6. If meter is left in bypass configuration, open drain valve (E) to prevent tube

damage caused by thermal expansion of the process liquid.