Gt1000 series – Brooks Instrument GT1000 User Manual

Page 9

Advertising
background image

1-3

Section 1 Introduction

Installation and Operation Manual
X-VA-GT1000-eng
Part Number: 541B078AAG
March, 2014

GT1000 Series

Table 1-1 Specifications - GT1000

Capacities and Pressure Drops

Capacities and Pressure Drops

Capacities and Pressure Drops

Capacities and Pressure Drops

Capacities and Pressure Drops

See Capacities and Pressure Drop Tables 1-3, 1-4 and 1-5

Flow

Flow

Flow

Flow

Flow Accur

Accur

Accur

Accur

Accuracy

acy

acy

acy

acy

Standard: ±2% Full Scale, Class 2.5 acc VDI/VDE
Optional: ±1% Full Scale, Class 1.6 acc VDI/VDE

Repeatability

Repeatability

Repeatability

Repeatability

Repeatability

≤ 0.5% Full Scale

Pressure Ratings/PED Categories

Pressure Ratings/PED Categories

Pressure Ratings/PED Categories

Pressure Ratings/PED Categories

Pressure Ratings/PED Categories

See Pressure Ratings/PED Categories Table 1-2 for maximum non-shock pressure

Scales

Scales

Scales

Scales

Scales

Single or dual detachable aluminum plate
Nominal Lengths: 127mm, 200mm and 250mm
Graduations: Choice of direct reading units, millimeter or percentage of maximum flow with factor tag

Ambient Temperature Limits

Ambient Temperature Limits

Ambient Temperature Limits

Ambient Temperature Limits

Ambient Temperature Limits

33 to 125°F (1°C to 52°C)

Operating Fluid Temperature Limits

Operating Fluid Temperature Limits

Operating Fluid Temperature Limits

Operating Fluid Temperature Limits

Operating Fluid Temperature Limits

Maximum: 250°F (121°C)

(Meter)

(Meter)

(Meter)

(Meter)

(Meter)

Minimum: 33°F (1°C)

Operating Fluid Temperature Limits

Operating Fluid Temperature Limits

Operating Fluid Temperature Limits

Operating Fluid Temperature Limits

Operating Fluid Temperature Limits

Maximum: 250°F (121°C) - Reed Switch

(Alarms)

(Alarms)

(Alarms)

(Alarms)

(Alarms)

Maximum: 167°F (75°C) - Inductive Switch
Minimum: 33°F (1°C) - Reed Switch/Inductive Switch

Dimensions

Dimensions

Dimensions

Dimensions

Dimensions

See Dimensions Figure 1-4

Materials of Construction

Materials of Construction

Materials of Construction

Materials of Construction

Materials of Construction

Metering Tube

Metering Tube

Metering Tube

Metering Tube

Metering Tube

Borosilicate glass

Window

Window

Window

Window

Window

Polycarbonate with UV inhibitor

Floats

Floats

Floats

Floats

Floats

Size 2 and 6: Sapphire, glass, Carboloy

®

, 316 stainless steel

Size 7: Glass, 316 stainless steel
Sizes 8-13: 316 stainless steel

Float Stops

Float Stops

Float Stops

Float Stops

Float Stops

Size 2, 6, 12 and 13: Teflon

®

Size 7, 8, 9 and 10: 316 stainless steel springs

Housing

Housing

Housing

Housing

Housing

Cast Aluminum with Epoxy paint

End Fittings

End Fittings

End Fittings

End Fittings

End Fittings

Brass or 316/316L stainless steel (1.4404)

O-rings

O-rings

O-rings

O-rings

O-rings

Viton

®

fluoroelastomers, Buna-N, Kalrez

®

perfluoroelastomers (Stainless steel body only), EPDM (Stainless steel body only)

Hardware

Hardware

Hardware

Hardware

Hardware

Stainless steel

Connections

Connections

Connections

Connections

Connections

Brass or stainless steel fittings: NPT or BSPT/RC female connections
Stainless steel fittings: 150 lbs. flanges per ANSI B 16.5

Connection Orientation

Connection Orientation

Connection Orientation

Connection Orientation

Connection Orientation

Vertical or horizontal on inlet and/or outlet

Certifications

Certifications

Certifications

Certifications

Certifications

Declaration of Compliance 2.1
International Calibration Certificate
Declaration of Compliance 2.1 Pressure Test
Hazardous Location Certificate
Material Certification to DIN 3.1
Declaration of Compliance 2.1 Oxygen Service

Table 1-2 GT1000 Pressure Ratings and PED Categories

Maximum Operating Pressure (PSIG/bar) at Fluid Temperature: Up to 250°F (121°C)

Meter

NPT

ANSI 150# RF

PED

Size

Threaded Connections

Flanged Connections

Category

2

500/34.5

240/16.5

SEP

6

450/31

240/16.5

SEP

7

300/20.7

240/16.5

SEP

8

250/17

240/16.5

SEP

9

200/13.8

200/13.8

SEP

10

175/12.1

175/12.1

SEP

12

100/6.9

100/6.9

See Note

13

75/5.2

75/5.2

See Note

Note: Size 12 and 13 do not conform to Pressure Equipment Directive 97/23/EC,
therefore cannot be sold or used in the EU/EFTA.

Advertising