Functions – Burkert Type 7616 User Manual

Page 13

Advertising
background image

13

Start-Up

Standard presetting:

Maximum stroke volume

Set the flow rate by turning the adjusting screw.

Fix the set flow rate with locking screw.

functions

8.3.

control diagram

8.3.1.

T0

T5

T4

T3

T2

T1

T6

ms

15

5

10

15

15

V1

V2

V3

T1

0

0

Control diagram

Fig. 6:

Via the parameters (T0 – T6), the times for the valve control are set.
By changing the interval times (T6 – T1), the frequency can be
changed.

times for maximum throughput:
While maintaining the times illustrated in the figure "Control
diagram" and an interval time of 0 ms, the maximum frequency
(15 Hz or 800 cycles/min) will be reached.
This corresponds to a throughput of approximately 4 ml/min with a
set maximum stroke volume.

recommended times for maximum accuracy:
For dosing with the highest accuracy (Tolerance and reproducibility)

≤ ± 2%), the following times are recommended:

T0

T5

T4

T3

T2

T1

T6

ms

50

50

30

50

50

T1

0

x

This corresponds to a throughput of approximately 1.15 ml/min with
a set maximum stroke volume.

english