Operating temperatures – Burkert Type S056 User Manual

Page 5

Advertising
background image

SENSOR FOR ELECTROMAGNETIC FLOW METER – MANUAL

5

ESENS_FULLBORE_05.doc

OPERATING TEMPERATURES

o

Shrewdness and precautions

NO

YES




































EBONITE LINING

PP LINING

PTFE LINING

Liquid Temp.

Ambient

Temp.

Liquid Temp.

Ambient

Temp.

Liquid Temp.

Ambient

Temp.

Min. Max Min. Max Min. Max Min. Max Min. Max

Min. Max

°

C 0 80 -5 60

0

60

0 60 -20 130 -10 60

° F

30

176

23

140

32

140

32

140

-4

266

14

140

LONG PIPE

3 X DN

2 X DN

ANTI VIBRATION JOINTS

GASKET THICKNESS + 4mm

For installations on long

pipes, please use the

anti vibration joints

For vertical installations is

preferable an ascending flow

Before tightening the nuts,

place both the flanges of pipe

and device as near together

as possible.

During operation, the pipe

must be either completely full

of liquid, or completely empty

Install the sensor at a min.

distance to bows in pipes and

hydraulic accessories

For vertical installations with

descending flow direction

contact the manufacturer

Avoid operating when the

pipe is partially empty

Avoid the installation next to

bows in pipes and respect

min. inlet and outlet sections

Avoid the approach of the

flange and counter flange

using the closing force of

the nuts

Avoid the installation of the

sensor in a long pipe,

without any support of the

same

LONG PIPE

Advertising
This manual is related to the following products: