Technical data

Fluid Components Intl

Document 06EN003327 Rev. -

7

FlexCOR

ä

Model CMF Series

Measurement of

Mass flow [lb/min / kg/s], volume flow [gpm / [l/s], fraction [%],

°Brix, density [kg/m

3

], temperature [°F,

°C]

Current output

Current

0-20 mA or 4-20 mA

Load

< 800 ohm

Time constant

0-30 s adjustable

Digital output

Frequency

0-10 kHz, 50% duty cycle

Time constant

0-30 s adjustable

Active

24 V d.c., 30 mA,

1 KW

£ R

load

£ 10 KW, short-circuit-protected

Passive

3-30 V d.c., max. 110 mA,

1 KW

£ R

load

£ 10 KW

Relay

Type

Change-over relay

Load

42 V / 2 A peak

Functions

Error level, error number, limit, direction

Digital input

11-30 V d.c. Ri = 13.6 K

W

Functionality

Start/hold/continue batch, 0-point adjust, reset totalizer 1/2, force output, freeze output

Galvanic isolation

All inputs and outputs are galvanically isolated, isolation voltage 500 volts

Cut-off

Low-flow

0-9.9% of maximum flow

Limit function

Mass flow, volume flow, fraction, density, sensor temperature

Totalizer

Two eight-digit counters for forward, net or reverse flow

Display

Background illumination with alphanumerical text, 3 × 20 characters to indicate flow rate, totalized
values, settings and faults.
Reverse flow indicated by negative sign

0-point adjustment

Manual via keypad or remote via digital input

Ambient temperature

Operation:

-20 to +50°C, max. rel. humidity 80% to 31°C decreasing to 50% at 40°C according to UL 3101

During storage:

-40 to +70°C (Humidity max. 95%)

Communication

Prepared for client mounted add-on modules

Enclosure

Material

Fiber glass-reinforced polyamide

Rating

IP 67 to IEC 529 and DIN 40050 (1 m w.g. for 30 min.)

Mechanical load

18-1000 Hz random, 3.17G rms, in all directions, to IEC 68-2-36

Supply voltage

24 V version

230 V version

Supply

24 V d.c./a.c., 50-60 Hz

115/230 V a.c., 50-60 Hz

Fluctuation

24 V d.c.,

-25 to 25%

+10 to -10%

24 V a.c.,

-16 to 25%

Power consumption

10 W

26 VA

Fuse

230 V version: T400 mA, T 250V (IEC 127) - Not to be changed by user
24 V version: T1A, T 250V (IEC 127) - Not to be changed by user

EMC performance

Emission

EN 50081-1 (Light industry)

Immunity

EN 50082-2 (Industry)

Namur

Within the value limits according to "Allgemeine Anforderung" with error criteria A in accordance with
NE 21

Evironment

Environmental conditions acc. to UL 3101:

Indoor use
Altitude up to 2000 m
POLLUTION DEGREE 2

Maintenance

The flowmeter has a built-in error log/pending menu which should be inspected on a regular basis

Mass Flowmeter Compact IP 67