Fluid Components International CMF Series Manual Technical Data User Manual

Page 6

Advertising
background image

Technical data

Fluid Components Intl

Document 06EN003327 Rev. -

11

FlexCOR

ä

Model CMF Series

2.3

Meter uncertainty

Display/frequency
and pulse output

Reference conditions (ISO 9104 and DIN/EN 29104)

Flow conditions

Fully developed flow profile

Temperature of medium

-4°F (20

°C) ± 2K

Ambient temperature

-4°F (20

°C) ± 2K

Liquid pressure

2

± 1 bar

Density

0.997 g/cm

3

Brix

40

°Brix

Supply voltage

Un

±1%

Warming-up time

30 min.

Cable length

16.4 F (5 m) between converter and sensor

Additions in the event of deviations from reference conditions

Current output

As pulse output

±(0.1% of actual flow +0.05% FSO)

Effect of ambient temperature

Display/frequency/pulse output: <

±0.003% / K act.

Current output: <

±0.005% / K act.

Effect of supply voltage

< 0.005% of measuring value on 1% alteration

· For flow > 5% of the sensors max. flow rate, the error can be read direct from the curve.
· For flow < 5% of the sensors max. flow rate, use the formula to calculate the error.
· The error curve is plotted from the formula:

E =

±

Ö

(0,10)

2

+

(

)

2

z

´ 100

qm

E = Error [%]
Z

= Zero point error [kg/h]

qm = Mass flow [kg/h]

Measuring pipe type

TRANSMITTER

Measuring pipe version

CMF - A

CMF - B

CMF - C

CMF - D

CMF - E

CMF - F

Number of measuring pipes

1

1

1

1

1

1

Mass flow:

·

Linearity error

% of rate

0.10

0.10

0.10

0.10

0.10

0.10

·

Repeatability error % of rate

0.05

0.05

0.05

0.05

0.05

0.05

·

Max. zero point error

[kg/h]

0.002

0.03

0.15

0.66

3.0

6.0

Density:
· Density error

[g/cm

3

]

0.001

0.0015

0.0015

0.0005

0.0005

0.0005

· Repeatability error

[g/cm

3

]

0.0002

0.0002

0.0002

0.0001

0.0001

0.0001

Temperature:
· Error

[

°C]

0.5

0.5

0.5

0.5

0.5

0.5

Brix:
· Error

[

°Brix]

0.6

1.2

0.4

0.2

0.2

0.2

CMF - A 1.0

40

65 kg/h

CMF - B 12

125

250 kg/h

CMF - C 50

500

1000 kg/h

CMF - D 280

2800

5600 kg/h

CMF - E 1250

12500

25000 kg/h

CMF - F 2600

26000

52000 kg/h

Advertising