Table 5-8. Components for 100 and 1000 Ohm Flowmeters

Input Board Replacement Procedure

Spare input boards MUST be factory pre-adjusted for the specific flowmeter and flow element sensing points. See
Figure 5-5 for the PWB parts placement.

1.

Turn the operating power OFF.

2.

Disconnect the flow element cable from the input board.

3.

Remove the four hold-down screws from the input board being replaced. Grasp the PWB and lift while gently
rocking the board from side to side.

4.

Position new PWB in the same orientation as the old PWB. Gently press down over the connector to seat it.

5.

Install the four hold-down screws in the new PWB.

6.

Reconnect the flow element cables to the PWB. Remove the "HTR+" and "HTR-" wires from the connectors
and continue with the next procedure.

7.

Perform a sensing point balancing procedure. Contact FCI customer service for the appropriate procedure.

8.

When the sensor balancing procedure is completed, seal all adjusted potentiometers.

9.

Turn the operating power OFF. Reconnect the heater circuit wires "HTR+" and "HTR-".

CONTROL BOARD

Adj. Pot.

100 Ohm

1000 Ohm

R8

4.99K, 1%

21.5K, 1%

Gain

20X

10X

INPUT BOARD

Adj. Pot.

100 Ohm

1000 Ohm

R1, R7

6.65K

Not Used

RN3

1K

10K

JUMPERS INSTALLED, INPUT BOARD

Dash No.

100 Ohm

1000 Ohm

OO1

N/A

J1, 2, 4, 6, 7, 8, 10, 12

OO2

N/A

J1, 2, 4, 6

OO3

J3, 5, 9, 11

N/A

OO4

J3, 5

N/A

JUMPERS INSTALLED, OUTPUT BOARD

4-20 mA Output

Jumper

Non Zero-Based

J1, 4

Zero-Based

J1, 5