FLUID COMPONENTS INTERNATIONAL LLC

ST50/ST75 PDA SOFTWARE

Chapter 6
Using the Configuration Software to Access the ST50/ST75 Diagnostic Functions


The Configuration Software gives access to three ST50/ST75 Diagnostic Functions. These functions help in troubleshooting the
ST50/ST75 operation, when a problem is suspected.

The ST50/ST75 gives the technician a window from which to analyze the processing of the flow signal information. This window
opens into 3 areas of the signal processing sequence. The first area is the signal as it comes out of the A/D in raw counts. The
second is as the signal is converted into ohms. And the third area is after the signal has been temperature corrected.

• Tap on the “Diagnostics” radio button of the main screen,

to open the ST50/ST75 Diagnostics/Troubleshooting
selection screen.

• Only one of the 3 areas of the diagnostics can be viewed

at a time.

• To look at a diagnostic data window, tap on the “Get”

button next to it.

• To exit Diagnostic mode tap on the “Return” button.