Number of static rinses, Number of tests, Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual

Page 146

Advertising
background image

Titra

Lab

®

TitraLab 860 and 865 Reference Manual

Page 146

Number of
static rinses

If a Sample Changer is in use, enter the desired number of static
rinses to be carried out before each sample run of a SAC sequence.

During a static rinse, the electrodes are dipped for a time (selectable
between 0 and 30:59 min:s,

see "Rinse time", page 189

) into a

beaker filled up with a rinse or conditioning solution.

These rinse beakers are located:

on the 1, 2 or 3 last positions of the SAC80 turntable,

on the RINSE 1, RINSE 2 or RINSE 3 dedicated rinse positions of
a SAC90 tray,

on the last 1 or 2 available positions of the SAC850 tray,

on the 1 or 2 dedicated rinse positions of the SAC950
reconditionning beakers extension.

Enter in:

Setup menu > Configuration

Range available:

0, 1, 2 for a SAC850 or SAC950.
If you have allowed the dynamic rinses to occur in previous beaker
except 1st in Rinse 2, then only one beaker will remain available for
static rinses (Rinse1 beaker).

Refer to "Dyn. rinse", page 90

.

0, 1, 2 or 3 for a SAC80/SAC90

Refer to User’s Guide of the sample changer used (part no.: D21T002
for a SAC90, D21T013 for a SAC80, D21T085 for a SAC850 or
SAC950).

Number of tests

This is the number of times you wish to repeat the method on the
same sample. The method will be repeated in a new beaker. If the
method is part of a coupled method the number entered here will not
be taken into account. It will be the number of tests entered in the
Coupled method parameters that will be used.

For a reagent calibration method, it is the number of beakers to pre-
pare for the calibration. The average value will be taken into account
when calculating the reagent concentration.

Enter in:

Edit method menu.
Edit reagent menu if Titre = Calibrate.

Range available:

1 to 99 (1 to 20 for a reagent calibration method).

Advertising