6 potentiometers, 7 excitation voltage select, 8 gain select switches – Cleveland Motion Controls Classic Series DIN Rail Amplifier Non-Isolated MWI-13466 REV CA User Manual

Page 12: Otentiometers, Xcitation, Oltage, Elect, Witches

Advertising
background image

C

LASSIC

S

ERIES

N

ON

-I

SOLATED

A

MPLIFIER

MAN-13466

R

EV

CA

P

AGE

12

OF

26

2.6 P

OTENTIOMETERS

In addition to the Gain and Zero adjustable Potentiometers visible on the front of the unit, there is an adjustment you
can make by removing the snap-off cover on the side of the amplifier. The adjustment is on the input printed circuit
board as shown in Figure 4. The following table provides you with a list of all the Potentiometers, where they are
located on the Classic Series DIN Rail input printed circuit board, and a description of their functions. To adjust
potentiometers, refer to Adjustment Tools 3.7.

Potentiometer

Reference

Designator

Location

Function

GAIN P2

Unit

Front

Provides 10:1 “vernier” adjustment of the variable gain amplifier. It is a
multi-turn potentiometer, with clockwise rotation causing an increase in
amplifier gain. When turned fully counter clockwise, the potentiometer will
cause the amplifier stage to provide the minimum gain of 1.8.

ZERO P1

Unit

Front

Provides a fine zero (offset) adjustment. It is a multi-turn potentiometer,
with clockwise rotation causing a positive shift in the analog outputs. It
should be set mid-way prior to setting the COARSE ZERO adjustment.

Coarse Zero

P5

Input, behind
access cover

Establishes the coarse zero of instrumentation amplifier. Because of the
ability to cause +/- Full scale (+/- F.S.) output shifts, it is important to
correctly follow the final set-up and calibration procedure so that premature
amplifier “clipping” is avoided

2.7 E

XCITATION

V

OLTAGE

S

ELECT

The Excitation Voltage is determined by the position of jumper switch J3. Refer to Figure 4 for Jumper-switch
locations. The jumper default setting is J3 (1-2) for 5.00V excitation Do not use 10V setting J3 (2-3) unless
explicitly permitted by the load cell electrical specifications. Promptly verify the excitation voltage after power-up
to avoid overdriving strain gages. Note that if there is no external load resistance, the voltage may rise to 6.4V, but
will immediately regulate at 5.00V when the load cells are connected.
Keep in mind that the strain gage based load cells can readily operate at less than rated voltage (with a
corresponding reduction in output signal). This fact is helpful in the event that a “10 V “ load cell exhibits an output
signal that is excessive for even the lowest amplifier gain.

2.8 G

AIN

S

ELECT

S

WITCHES

The group of jumper-switches that control the fixed gain of the Instrumentation Amplifier (IA) are located closest to
the front left edge of the Input card. (Refer to Figure 4) The lowest gain (Av = 5) occurs when all of the switches
are in the 1-2 position. As switches are moved “away” (into position 2-3) from left to right, the gain progressively
increases as described in following table:

Jumper Switches

J10 J9 J7

Voltage Gain

1-2 1-2 1-2

5

2-3 1-2 1-2

25

2-3 2-3 1-2

125

2-3 2-3 2-3

620

When changing the internal Jumper-switch settings, it is always advisable to change the settings
with the 24 VDC power removed. If this is not possible, it becomes particularly important to use
a non-conductive tool to alter switch positions.

Make sure that Jumper- switch settings are fully in position to avoid accidentally leaving a switch
in an “in-between” state.

The total gain range of the amplifier is the product of the IA and variable amplifier gains
(variable).
Using the lowest gain switch settings, the net gain is 5 x (1.8 to 18) or 9 to 90
With the other gain combinations available, a 10 volt output can be produced with input voltages
ranging from between 0.002 to 0.50 volts.

Advertising