Model T675 Flow Totalizer

Technical Information

Pg. 23

3.3.1.6 – Printer Type
Allows selection of a specific printer output format or disables the printer output. Depress the corresponding
number in left column to access that menu item. Any selection automatically disables the remaining menu items. If
the printer output has been disabled the Print softkey will not be displayed.

Example 1: Disable printer output: <1>.
Example 2: Enable OBC: <6>.

Note 1: The time that delivery information is available to the printer after flow discontinues is

programmed in the Print timeout menu.

Note 2: Menu item 4 is intended for specific printers that require print coordinates to be

defined such as Epson or Zebra brands. Predefined with titles outputs at the
specified coordinate the titles associated with the values. The coordinates are
programmed in the Printer format 1/2 and 2/2 menus.

Note 3

************* Printer type *************

: When printer type 5 is selected, the printer format coordinates are copied from the

default.

Current setting = 1


1 – Disabled
2 – SP824
3 – SP834
4 – Predefined with titles (online ck)
5 – SP84x
6 – OBC
7 – ALC TELXON


Enter setting (ESC to exit)?

3.3.1.7 – System alarm log
Displays a chronological log of 16 system alarms divided into 2 groups of 10 events, Alarm Log page 1(newest)
and Alarm Log page 2 (oldest). Each entry contains the date, time, description of the failure and value. The
Alarm Log page 1 contains the most recent events listed top to bottom. After more than 10 events the oldest
listing on page 1 will transfer to the Alarm Log page 2. The Empty designation indicates that less than 10 events
have been registered on that page.
Note:

If the current alarm is the same as the previous alarm in the table it will be classified as a continuation of the

previous alarm and will not be added to the table as a separate event. This is to prevent the same alarm
condition from overflowing the table.

********** Alarm log, page 01 **********

010 21-Mar-07 18:48 –Coil open
009 08-Sep-07 17:20 –RTD short
008 12-May-07 16:16 –Low power
007 16-Feb-07 10:51 –Tloop short
006 13-Oct-07 06:35 –Coil open
005 11-Jul-06 14:25 –Tloop open
004 15-Apr-06 09:42 –Low batt
003 03-Feb-06 07:33 –RTD open
002 31-Jan-06 01:15 –Coil short
001 11-Nov-05 20:30 –Low power

Press any key to continue