Upgrade firmware, Luminex sd system firmware upgrade, Close any open applications – Luminex IS 2.1_2.2 to xPONENT 3.1 Rev 2 Upgrade Instructions User Manual

Page 11: Turn off the power to the sd, xyp, and analyzer, Turn power back on to the sd system

Advertising
background image

Luminex IS Versions 2.1 and 2.2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use

7

To install the xCONVERT Data Conversion Utility:

1. Click Install xPONENT file conversion utility. A File Down-

load – Security Warning dialog box opens.

2. Click Run. An Internet Explorer – Security Warning dialog

box opens.

3. Click Run. An xCONVERT Setup dialog box opens (.NET

Framework 2.0).

Note: Steps 3 and 4 may not
occur in all installations.

4. Click Accept. An xCONVERT Setup dialog box opens (for

SQL Server 2005 Express Edition).

5. Click Accept. An xCONVERT Setup dialog box opens (Visual

C++ Runtime Libraries (x86).

6. Click Install. A series of installation dialog boxes open as the

software installs. This may take several minutes.

7. The xCONVERT Setup Wizard opens. Click Next to continue.
8. Select an installation folder for the software. Click Next to

continue.

9. The Confirm Installation dialog box opens. Click Next to

continue.

10. An Installation Complete dialog box opens. Click Close to

complete the installation

.

Upgrade Firmware

Upgrade Firmware

The Luminex SD

system, analyzer, and XYP instrument use

firmware to perform internal control functions. You must perform a
firmware upgrade for each of these instruments when you upgrade
the software. You must upgrade the firmware in the following order:

SD (if present)

LX100/200

XYP

If you are using an existing computer, perform the Luminex software
update before upgrading the firmware. See "Install Luminex
xPONENT 3.1 SP1 Software".

Luminex SD System

Firmware Upgrade

To upgrade the Luminex SD system firmware:

1. Close any open applications.
2. Turn off the power to the SD, XYP, and analyzer.
3. Remove the serial cable from the Luminex XYP instrument and

plug it into P1 on the back of the SD system.

4. Turn power back on to the SD system.
5. Click Start > Programs > Luminex > LXR > LxFirmware on

the PC desktop. The LxFirmware Update Tool dialog box
opens.

Advertising