8 main display, Configuration settings, Not currently used – Meriam 4150X ATEX I.S. Certified HART Communicator User Manual

Page 12

Advertising
background image

8

Main Display

The Main display provides access to Online Setup, Offline Menu and to
4150 / 4150X configuration settings via the right hand Soft Key (Cfg).

1. Online Setup is described above in Turning On the 4150 / 4150X.

2. Offline Menu is selected to provide access to Offline Utilities

such as List / Show DOFs, List / Edit Configurations, and
Create Configurations. The number of DOFs installed (HART
device profiles) and the number of device Configurations stored
are also available on Offline Utilities screen. See the Offline
section of this manual for more information.


3. Cfg Soft Key is selected to configure 4150 / 4150X settings and

view unit information such as Model, Serial, and Firmware
numbers, DOFs installed and Configurations stored.


Configuration Settings


1 Users

Not currently used

2 Applications

Various applications are available here including Digital Poll (polling device addresses 0 – 15) and
other user and factory applications.

3

Lockouts

The HART Communicator can be programmed to lockout certain standard functions that a supervisor may
wish to control. The functions included are used to set up the 4150 / 4150X for normal use, change online
HART device configurations, create offline HART device configurations, update the installed DOFs list
(HART device profiles), or to update the Unit’s operating firmware. For example, locking out the Configs
feature makes the unit a read only HART communicator while locking out DOFs and Firmware prevents
unauthorized updates. The lockout can be limited to individual Lockout Details items or all Lockout Details
items can be locked out at one time.

To enable the lockout feature, press the “3” key or move the selection cursor

to the Lockouts menu

option on the Configuration screen (see below left). Choose the Lockouts selection by pressing the Chng
soft key. The 4150 / 4150X will enter the Lockout Enable / Disable Screen. The three choices and current
status information are provided (see below center):









Configuration Screen

Lockout Enable / Disable Screen

Lockout View Details Screen

Configuration #

Current Settings

1

Users

2

Applications

3

Lockouts

4

Clock/Timers

5

Other



Up Down Chng Back

Communicator

#

1

Online Setup

2

Offline Menu


Device 0: *On-Line*
Rosemount
3051C
Fld Dev Rev 03
Dof Name 26060309

Up Down Select Cfg

Configuration #

Lockout Details
1 All

:

Disabled

2 Setting :

Enabled

3 DOFs

:

Disabled

4 Configs :

Disabled

5 Firmware : Disabled

6 PV Prompt: Enabled

Up Down Chng Back

Configuration #

Lockouts
1 Lockout: Disable
2 PV Prompt: Disable
3 View Details

Up Down Chng Back

Configuration #

Current Settings

1 Users
2 Applications
3 Lockouts
4 Clock/Timers
5 Other

Up Down Chng Back

Advertising