P5 – setpoint options – Meriam M2110L SMART LEVEL GAUGE User Manual

Page 17

Advertising
background image

9R143 Dated 3/2011

15

P5 – SETPOINT OPTIONS

This register defines the output action for SetPoint and Current Loop models. Note that if an output is disabled by
this register, the SetPoint value in its corresponding register (P6, P7) has no impact. Also note that a SetPoint value
(P6, P7) may be 0.0, which is a viable output setting; thus, in order to disable control action, register P5 should be
used to define the desired action.

In Measure Mode, the indicators “SET1” and/or “SET2” will illuminate when the corresponding relay is energized.
The “4-20mA” indicator will illuminate when the Current Loop output is enabled. In Program Mode, the
indicators “SET1”, “SET2”, and “4-20mA” will illuminate as appropriate to assist when scrolling through the
register’s choices. Note that if P5 is set for active, the outputs will continue to update even in Program Mode,
according to the currently programmed values.

QUICK REFERENCE: P5, SETPOINT OPTIONS

Px

Name

Description

Value Range

Notes

SetPoint model:

Defines which relay outputs are active.

0 = Disabled.
1 = SET1 only.
2 = SET2 only.
3 = Both enabled.

P5 SetPoint

Options

Current Loop model:

Defines status of the 4-20 mA output.

0 = 4-20 disabled
1 = 4-20 enabled

Not found on
Battery model.

STEP BY STEP: P5, SETPOINT OPTIONS

Step Action

Display

1

Gauge should be in the normal Measure Mode. Normal

Pressure

Display.

Press the PRGM/Enter

→ key.

2

If the lockout is active, the gauge will now prompt for entry of
the lockout code. Refer to “STEP BY STEP: ENTERING A
LOCKOUT CODE (When Prompted)” on page 21.

The PRGM annunciator shows at the top of the
display, and the display shows the register name
P0”.

Press the Up

↑ or Down↓ keys repeatedly to scroll

to the desired register.

3

The Zero function (Up

and Down↓) will reset the scroll to P0

and allow selection to continue.

Display shows “P5”.

4

Press PRGM/Enter

→ to open the register.

The display shows the current value.

Press the Up

↑ or Down↓ keys repeatedly to scroll

to the desired register value.

5

The Zero function (Up

and Down↓) will reset the scroll to

default and allow editing to continue.

All available choices are scrolled.

6

Press the PRGM/Enter

→ key to accept the value.

The value is accepted, and the register is closed.
Display shows “P5”.

7

Press the Backspace

← key to activate the register

setting and return to Measure Mode.

Normal Pressure Display.

Notes

1. After steps 2 and 3, the Backspace

← key will abort the process and return to Measure Mode.

2. After steps 4 and 5, the Backspace

← key will abort the edit and close the register.

3. After step 6, the PRGM/Enter

→ key will again open the register for edit.

4. During any programming operation, if there is no keypad activity for approximately 1 minute, the

operation is aborted and the gauge returns to Measure Mode unchanged.

SetPoint & Current Loop Models Only

Advertising