Range limits and indication – Meriam M2110L SMART LEVEL GAUGE User Manual

Page 31

Advertising
background image

9R143 Dated 3/2011

29

RANGE LIMITS AND INDICATION




Sensor Hard

Soft

Certified

Soft

Hard

Type

Under Under

Range

Over

Over

Range Range

(F.S.)

Range

Range

DN

Under

-20%

1

-20 to 0%

0 to 100%

100 to

120%

Over

120%

D

I

Under

-20%

2

-20 to 0%

0 to 100%

100 to

120%

Over

120%

G

I

Under

-20%

3

-20

3

to

0%

0 to 100%

100 to

120%

Over

120%

C

I

NA

4

NA

4

0 to 100%

100 to

120%

Over

120%

A

I

NA

5

NA

5

0 to 100%

100 to

120%

Over

120%

Notes:

1

DN units will measure to -20% (or -10 PSID, whichever is greater

pressure relative to absolute zero) and +120% of full scale pressure.
Calibration certification is for 0 – F.S. (Full Scale Only).

2

D

I

units will measure to -20% (or -150 PSI, whichever is greater

pressure relative to absolute zero) and +120% of full scale pressure.
Calibration certification is for 0 – F.S. (Full Scale Only).

3

G

I

units will measure to -20% (or -10 PSIG, whichever is greater

pressure relative to absolute zero) and +120% of full scale pressure.
Calibration certification is for 0 – F.S. (Full Scale Only).

4

C

I

units will measure to negative barometric pressure (typically -14.7

PSIG) and to +120% of full scale pressure. Calibration certification is for
-14.5 PSIG to F.S. (Full Scale Only).

5

A

I

units will measure to +120% of full scale pressure. Calibration

certification is for 0 PSIA to F.S. (Full Scale Only).

Hard Over & Under:
Display will read "OP" (Under -20% or Over 120%).

Soft Over & Under:
Display will still read out at (± 20%).





Advertising