19. Outline Dimensions

20. Panel Mounting

20. Panel Mounting

1. Make a ¼ DIN panel cutout per drawing.

2. Remove the two 6-32 socket head screws in

the grooves at the rear of the gauge.

3. Slide the panel mount jacks out of the

groove.

4. Insert the gauge through the front of the

panel.

5. From the rear, insert the panel mount jacks in

the grooves on the side of the gauge, and
slide them firmly against the panel.

6. Replace the two 6-32 socket head screws in

the grooves behind the panel mount jacks.

7. Tighten the panel mount jack against the

panel with the two socket head screws.

8. Refer to the note on page 28 when making

pressure

connections.

3.60" +/- 0.010

(92)

( ) DIMENSIONS IN MILLIMETERS

3.60" +/- 0.010

(92)

PANEL MOUNTING CUTOUT DIMENSIONS