4. Programmable Register Overview

All M2110P Smart Pressure Gauges have programmable registers that allow the gauge to be configured to fit
the pressure measurement application. Programmable registers are numbered P0 through P8. Each register
controls a specific aspect of the gauge’s performance.

A description of all programmable registers, and step by step instructions for their use, is included below. Each
register is found on all M2110 series models, and performs the same function described for all models, except
as noted otherwise.

Px

Name

Description

Value Range

Notes

P0

Lockout Code

Lockout for security.

00 to 99

00 = Disabled.

P1

Timeout

Automatic shutoff in minutes of keypad
inactivity.

0 (disabled), 1, 2, 5,
10, 15, 25

Battery model only.

P2

Damp Rate

Exponential damping time in seconds.

0.1, 0.2, 0.5, 1, 2, 5,
10, 15, 25, 50

0.1 = No Damping.

P3

Density of Water
Reference
Temperature

Sets reference temperature for water
density when gauge is set for water
column units.

20°C, 4°C, or 60°F

P4

User Units
Full Scale

Enables gauge to display units other
than factory defined Engineering Units.

0 to 9,999 (Keypad)
0 to 19,999 (Serial)

User defined value.

SetPoint Model: defines which relay
outputs are active.

0 = Disabled.
1 = SET1 only.
2 = SET2 only.
3 = Both enabled.

P5

SetPoint Options

Current Loop Model: defines the status
of the 4 - 20 mA output.

0 = 4-20 disabled
1 = 4-20 enabled

Not found on
Battery model.

P6

SET1

Controls SET1 relay or 4.00 mA value.

–20% to +120% FS

User defined value.

P7

SET2

Controls SET2 relay or 20.0 mA value.

–20% to +120% FS

User defined value.

P8

Deadband

Sets the amount of deadband in percent
of full scale for relays.

0 (disabled), 0.1,
0.2, 0.5, 1, 2, 5, 10%

SetPoint model
only.