Meriam M4 Series Precision Calibrator Data Logger Manual User Manual

Page 18

Advertising
background image

ZM4 Instruction Manual 9R87-B

June 2012

18

Damping Setup

Figure 13: Damping Setup Display


Use the Type key to change the Damping Type. The options are a follows:

1) None. The measured value is not filtered or damped.
2) Exp(exponential), which uses a traditional damping algorithm with

response delays associated with the Damping Rate.

3) Smart, which employs a smart damping algorithm to produce and

instant response to a step change. After the step change it uses the
Damping Rate to damp the measurement exactly as in exponential
damping.


Use the arrow keys and the Adaptive Interface Wheel to set the Damping
Rate in seconds. The value can be adjusted between 0 and 32 seconds.

Press the Exit key to return to the main Pressure Sensor Setup menu.