57-649 Jupiter Magnetostrictive Transmitter - F

OUNDATION

fieldbus™

23

5.1.2

Status Messages

Display Message

Action

Comment

OK

None

Normal operating mode

Initial

None

Shown at power-up during
self check

Default Cal

Factory set default calibration parameters

Consult Factory

are in use, level reading may be inaccurate

Lo Temp

Present temperature in electronics

Transmitter may need to be moved

compartment is below -40° C

to ensure temperature is within
specification

Hi Temp

Present temperature in electronics

Transmitter may need to be moved

compartment is above +80° C

to ensure temperature is within
specification

Float 2

No level signal detected from float 2

Make sure 2 floats are being used,

Fail

are not damaged, and within
measuring range

Float 1

No level signal detected from float 1

Make sure float is not damaged

Fail

and within measurement range

No Signal

No signal detected from any float

Make sure float is not damaged
and within measurement range

Snsr Brd Fail

No signal from probe, bad board or

Check electrical connections

connection problem

to probe

DfltParm

Internal non-volatile, parameters have

Consult Factory

been defaulted