MEASUREMENT WITH SET PRESSURE

The UA-1020 is designed to detect the pulse and to inflate the cuff to a systolic

pressure level automatically. If re-inflation occurs repeatedly, use the following

methods.
During the blood pressure measurement, re-inflation may occur. A fixed pressure

value can be set to avoid re-inflation.
1. At step 6 of the clock adjustment procedure (see page

E-10), press the

SET

button to go to the pressure setting

mode. The current setting blinks.

2. Press the or button to select a pressure value

about 30 mmHg or more above your expected systolic

pressure from the following.

AUTO

: Automatic pressurization (default value)

180

: Pressure value of 180 mmHg (fixed)

210

: Pressure value of 210 mmHg (fixed)

240

: Pressure value of 240 mmHg (fixed)

3. Press the

SET

button to go to the TriCheck setting mode.

Press the

START

button to turn the device off. After three minutes of

non-operation, the device will turn off automatically. The next measurement

will be performed with the new pressure value.

Turn off

TriCheck

setting mode

or

START

Refer to previous page

for measurement

Press and hold the button

At heart level

Release the button

at the desired

systolic pressure

MEASUREMENT WITH DESIRED SYSTOLIC PRESSURE

Use this method when re-inflation occurs repeatedly even if the pressure value is

set to

240

in the procedure above or when the results are not displayed even if

the pressure decreases to 20 mmHg or less.
1. Place the cuff on the arm (preferably the left arm).
2. Press and hold the

START

button until a number

about 30 to 40 mmHg higher than your expected

systolic pressure appears.

3. When the desired number is reached, release the

START

button to start measurement. Continue to

measure your blood pressure as described on the

previous page.

Turn off

TriCheck

setting mode

or

START

Refer to previous page

for measurement

Press and hold the button

At heart level

Release the button

at the desired

systolic pressure