E-27

A&D Medical (Toll-Free): 1-888-726-9966

EMC TABLE

Medical Electrical Equipment needs special precautions regarding EMC and needs to be

installed and put into service according to the EMC information provided in the following.

Portable and mobile RF communication equipment (e.g. cell phones) can affect Medical

Electrical Equipment. The use of accessories and cables other than those specified may result

in increased emissions or decreased immunity of the unit.

Guidance and manufacturer’s declaration – electromagnetic emissions

The A&D unit is intended for use in the electromagnetic environment specified below.
The customer or the user of the A&D unit should assure that it is used in such an
environment.
Emissions test

Compliance Electromagnetic environment – guidance

RF emissions

CISPR 11

Group 1

The A&D unit uses RF energy only for its

internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.

RF emissions
CISPR 11

Class B

The A&D unit is suitable for use in all
establishments,

including

domestic

establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for

domestic purposes.

Harmonic emissions
IEC 61000-3-2

Class A

Voltage fluctuations /

flicker emissions
IEC 61000-3-3

Complies

Recommended separation distances between portable and mobile RF communications
equipment and the A&D unit
The A&D unit is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the A&D unit can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the A&D unit as recommended

below, according to the maximum output power of the communications equipment.

Rated maximum

output

power of transmitter

W

Separation distance according to frequency of transmitter

m

150 kHz to 80 MHz

d = 1.2

P

80 MHz to 800
MHz

d = 1.2

P

800 MHz to 2.5 GHz

d = 2.3

P

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to
the frequency of the transmitter, where p is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency

range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is

affected by absorption and reflection from structures, objects and people.