Emc table – A&D Premier/TriCheck Blood Pressure MOnitor UA-1020 User Manual

Page 29

Advertising
background image

E-28

In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991

EMC TABLE

Guidance and manufacturer’s declaration – electromagnetic immunity

The A&D unit is intended for use in the electromagnetic environment specified below. The
customer or the user of the A&D unit should assure that it is used in such an environment.

Immunity

test

IEC 60601

test level

Compliance

level

Electromagnetic environment –

guidance

Portable and mobile RF

communications equipment should
be used no closer to any part of
the A&D unit, including cables,
than the recommended separation
distance calculated from the

equation applicable to the
frequency of the transmitter.

Recommended separation distance:

Conducted RF
IEC 61000-4-6

3 V

rms

150 kHz to 80 MHz

3 V

rms

d = 1.2

P

Radiated RF
IEC 61000-4-3

3 V/m
80 MHz to 2,5 GHz

3 V/m

d = 1.2

P

80 MHz to 800 MHz

d = 2.3

P

800 MHz to 2,5 GHz

where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation

distance in metres (m).

Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,

a

should be less than the compliance

level in each frequency range.

b

Interference may occur in
the vicinity of equipment
marked with the following
symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is

affected by absorption and reflection from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio

(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM

radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in
the location in which the A&D unit is used exceeds the applicable RF compliance
level above, the A&D unit should be observed to verify normal operation. If abnormal

performance is observed, additional measures may be necessary, such as re-
orienting or relocating the A&D unit.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.