Perma Pure GASS-2040 User Manual

Page 21

Advertising
background image

21

GASS-2040 User Manual | Original Instructions

APPENDIX A

Specifications

MAXIMUM SAMPLE FLOW RATE:

0 TO 10 LITER/MIN

MAXIMUM INLET SAMPLE TEMPERATURE:

250

°F W/SST FILTER

230

°F W/KYNAR FILTER

MAXIMUM GAS SAMPLE WATER VAPOR CONTENT:

70%

OUTLET SAMPLE DEW POINT:

26

°F (-4°C) at 10 L/MIN

(With GS2040-PD200KA dryer)

12

°F (-12°C) at 5 L/MIN

-12

°F (-25°C) at 2 L/MIN

36

°F (2°C) at 10 L/MIN

(With GS2040-PD100KA dryer)

26

°F (-4°C) at 5 L/MIN

6

°F (-15°C) at 2 L/MIN

(With GS2040-50KA dryer)

42

°F(5°C) at 10 L/MIN

SOLUBLE GAS REMOVAL RATES:

NO, NO2, SO2 0% loss
CO, CO2

0% loss

H2S, HCl

0% loss

MAXIMUM GAS SAMPLE INLET PRESSURE:

10 PSIG

MINIMUM GAS SAMPLE INLET PRESSURE:

- WITHOUT PURGE EDUCTOR OPTION:

5" H2O VACUUM

- WITH PURGE EDUCTOR OPTION:

10” H2O VACUUM

GAS SAMPLE INLET FITTINGS:

1/4" or 3/8" TUBING FITTINGS

GAS SAMPLE OUTLET FITTINGS:

1/4" or 3/8" TUBING FITTINGS

AIR REQUIREMENTS:

PURGE AIR -40

°C DEW POINT

MAXIMUM ONE (1) CFM

-OPTIONAL PURGE EDUCTOR

MAXIMUM THREE (3) CFM

ELECTRICAL REQUIREMENTS:

950 Watts @ 115V ±10% or 230V ±10%,
50/60Hz, 10A/5A

FUSE:

- FUSE BLOCK

1.25” X 0.25” GLASS FUSE

- INLINE FUSE HOLDER (PLC)

5mm X 20mm 1.0A FAST-BLOW FUSE

ENCLOSURE:

NEMA 4X, STAINLESS STEEL

DIMENSIONS:

20”W x 40”H x 10”DEEP

OPERATING ENVIRONMENT:

-20

°C TO 40°C AMBIENT TEMP.

0 TO 95% R.H.
Altitude - up to 2000m

Advertising