Troubleshooting guide

The SDS-Series is a very simple component of the sample conditioning system with no electricity,

heat or moving parts to fail. Problems can arise if there is a failure upstream which causes liquid
carryover or a high pressure or vacuum spike in the line. If these issues are suspected, please call

the factory. We have a dedicated technical support and repair department to help in trouble-
shooting and replacement parts.

Symptom Diagnostic

Fix

No purge flow

Check pressure of the
instrument air.

Supply correct pressure, 30-150
psig.


Restricted sample flow


Bypass the SDS system and
observe any changes.



If the restriction disappears, the
dryers in the SDS have failed and

may need replacement; call
factory at 732-244-0010 ext-145

for assistance.

If no change is observed, check
pump and other flow path

components.

No drying

Check instrument air pressure
and quality.


Check for flow at exhaust

point.

Check inlet for condensation in
sample line.

Oil-free instrument air; 30-150
psig.


10-12 lpm for SDS-D201

20-24 lpm for SDS-D202

Call factory at 732-244-0010 ext-
145 for instructions on drying out
system.

Replacement parts

3DYK-001

Replacement membrane dryer

3RPA-001

Replacement instrument air regulator