Perma Pure Modular Board User Manual

Page 14

Advertising
background image


Move switch #5 to the ‘OFF’ position of the DIP Switch labeled SW1 to exit
board configuration and return to operation. The updates are automatically saved.



Single Channel Cooler with communications
DIP Switch 1, 2 and 3 are OFF, OFF and OFF.
Menu

Item

Display Range


Set Point 1

SP1

-10 to 25 °C

Ready

1 rd1

-10

to

25

°C

Relay

Control

rLY

Reset

or

Latch

Modbus ID

I d

1 to 255

Set Point 1 is the temperature the cooler will try to maintain the impinger at
during operation.


Ready 1 is the temperature that impinger must be below to turn on the relay and
external equipment (i.e. sample pump)

Relay Control determines if the control relays reset automatically or latch on
alarm. Choose “Reset”, displayed as “rSt”, to have the relays reset automatically.
Choose “Latch”, displayed as “LAt”, to have the relays latch on alarm. An
external switch is required to reset the alarms.

Modbus ID is the address used in serial communication to the unit.



Single Channel Cooler 2

nd

channel

DIP Switch 1, 2 and 3 are ON, OFF and OFF.
Menu

Item

Display Range


Set Point 1

SP1

-10 to 25 °C

Ready

1 rd1

-10

to

25

°C

Relay

Control

rLY

Reset

or

Latch

Set Point 1 is the temperature the cooler will try to maintain the impinger at
during operation.


Ready 1 is the temperature that impinger must be below to turn on the relay and
external equipment (i.e. sample pump)

Relay Control determines if the control relays reset automatically or latch on
alarm. Choose “Reset”, displayed as “rSt”, to have the relays reset automatically.
Choose “Latch”, displayed as “LAt”, to have the relays latch on alarm. An
external switch is required to reset the alarms.